atlas-ametek logoSkip to Main Content
No options found

Performance Checks With Reference Materials

This Atlas Weathering Blog Post explains how weathering instrument precision varies by design and how performance checks using Blue Wool, Polystyrene Chip, or ORWET references ensure consistent degradation behavior through Delta E color change analysis over time and exposure.

fictional performance level indicator between 90 to 100 %

Differences Between Instrument Types

Weathering instruments face the challenge to simulate irradiance, spectrum, air temperature, surface temperature, and relative humidity simultaneously with sufficient precision over the entire test area. Filtered xenon-arc devices come in various forms, including flatbed and rotating rack designs. Their sample areas range from 500 to 12,000 cm², and control technologies such as sensor type and sensor location differ considerably. Consequently, the precision, i.e. repeatability and reproducibility, depends strongly on the type of instrument.

Performance Checks of New Weathering Devices

Whenever you install a new weathering device in your laboratory, one of your first actions you should take would be conducting your typical weathering test with a standard reference material. Conducting such an instrument performance check will show you if your new instrument performs on the same level as your existing device(s). The new instrument should demonstrate comparable, ideally identical performance. In other words, cause similar or (hardly achieved) identical degradation of the tested reference material.

Standard Reference Materials

Blue Wool Scale, SAE Polystyrene Chips (PS), and ORWET standard reference materials are proven useful for conducting performance checks. While Blue Wool reference materials are extensively utilized for interior testing, particularly for color light fastness and hot light fastness evaluations, PS chips and ORWET plates serve both interior and exterior weathering tests. PS chips are particularly useful for gauging responses against the UV cut-on wavelength and temperature, thereby indicating potential discrepancies in Daylight filter quality. On the other hand, ORWET plates are useful as their coating reacts to UV, visible radiation, and temperature and is additionally sensitive towards moisture.

Delta E Comparison by Time and by UV Radiant Exposure

One of the easiest analytical methods for above reference materials is the measurement of their color change (Delta E) over time under your preferred test conditions. The appropriate measurement intervals should be chosen according to the test duration.

 

Best graphics for analysis include Delta E plotted against test time and/or Delta E plotted against UV radiant exposure. Below is an example of a PS-check conducted within two Ci5000 Weather-Ometer® instruments installed at separate locations running ASTM D7869.

Performance check of two Ci5000 Weather-Ometers running ASTM D7869 conditions using ORWET reference material on the middle section of the specimen rack
Performance check of two Ci5000 Weather-Ometers running ASTM D7869 conditions using ORWET reference material on the middle section of the specimen rack
Dr. Oliver D. Rahäuser
About Dr. Oliver D. Rahäuser
Senior Product Manager

Dr. Oliver Rahaeuser user studied Chemistry in Wuerzburg, Germany. He received his Diploma in Inorganic Chemistry and Material Science, followed by his Ph.D. Oliver joined Atlas MTT in 2006 functioning as Product Manager for the SUNTEST® and Xenotest® accelerated weathering testing instrument product lines. Since 2013, he added Product Management of Atlas Custom Systems, which primarily focuses on tailored solar simulator solutions for small and large test chambers for component and product testing. Dr. Oliver Rahaeuser is a member of the German Colour Fastness Committee, and the DKE committees on PV and environmental testing of electronics. Furthermore, he consults Atlas clients on accelerated aging testing and test method development.