ICH Testing of Pharmaceuticals – Part 1: Exposure Conditions
December 13, 2021
Author:Dr. Oliver D. Rahäuser
The Atlas Weathering Blog Post explains ICH Q1B photostability requirements, SUNTEST® Option 1 UV/visible exposure behavior, and how controlled xenon testing ensures compliant, efficient confirmatory and forced‑degradation studies.
ICH Q1B Requirements for Photostability Testing
The FDA (American Food & Drug Administration) requires that all new drugs be tested for shelf life by exposing drugs to simulated sunlight. The international ICH guidelines define appropriate light sources and provides test guidelines. The guidelines Q1B state that drugs must be exposed to a minimum of 1.2 million lux-hours of visible light and a minimum of 200 W·h/m2 UVA light. These tests are also referred to as Confirmatory Studies.
Further, the ICH guidelines request so called Forced Degradation studies. The test objective here is to force the drug substance to degrade to a point that reveals potential structural weaknesses, which can be addressed either by chemical modifications or packaging. Here, the test device may run for longer duration and/or is using the harsher D65 light condition of simulated outdoor sunlight.
SUNTEST (Option 1) Photostability Chambers and the Question About UV Over Exposure
What is meant by this statement and more importantly, is it acceptable to use? As a matter of fact, when a SUNTEST achieves the minimum visible light dose, the minimum UV dose will be exceeded by a factor of approximately 2.4.
Since the ICH guidelines have only defined minimum values for both the Ultraviolet (no less than 200 W·h/m2) and the Visible (no less than 1.2 million lux-hours), it is perfectly acceptable to expose specimens to the longest exposure of the two. Most Atlas clients using this method typically apply at least 2-3x the minimum exposure requirements, while some may test up to 5x or even 10x. However, if “overexposure” is a concern of yours, simply begin your test with two replicate samples, pulling one replicate from the test when it has reached the UV minimum exposure. The remaining replicate can then be exposed until the minimum visible light dose is achieved.
Test Duration Inside SUNTEST
The two biggest benefits of a SUNTEST are great speed of testing and superior reliability of the solar simulation compared to Option 2 type photostability chambers.
ICH Q1B Testing
SUNTEST CPS
SUNTEST XLS
SUNTEST XXL-FD
Shortest time required for 200 Wh/m2
3.0 h
4.0 h
4.0 h
Shortest time required for 1.2 million luxh
6.9 h
9.2 h
9.2 h
About Dr. Oliver D. Rahäuser
Senior Product Manager
Dr. Oliver Rahaeuser user studied Chemistry in Wuerzburg, Germany. He received his Diploma in Inorganic Chemistry and Material Science, followed by his Ph.D. Oliver joined Atlas MTT in 2006 functioning as Product Manager for the SUNTEST® and Xenotest® accelerated weathering testing instrument product lines. Since 2013, he added Product Management of Atlas Custom Systems, which primarily focuses on tailored solar simulator solutions for small and large test chambers for component and product testing. Dr. Oliver Rahaeuser is a member of the German Colour Fastness Committee, and the DKE committees on PV and environmental testing of electronics. Furthermore, he consults Atlas clients on accelerated aging testing and test method development.